HA-RO constrict

Bild HA-RO Constrict - VE-Wasseraufbereitungsanlage

The HA-RO constrict is a high-quality VE water treatment system that is housed in a completely enclosed and low-noise system cabinet is installed. The reverse osmosis process ensures water qualities according to ASTM Type III and Type IV as well as DIN ISO 15883 reached. The pure water quality is 5 - 15 µS/cm, depending on the raw water quality.

The system is controlled via our proven ECO control unit with two-line operating display. This enables the intuitive operation of the system with a minimum noise level. The processing of measured values thus runs completely via the microprocessor control, so that separate field regulation and control devices for level, conductivity, temperature and pressure are no longer required. The system is located in a compact system cabinet and can be integrated into any laboratory and hospital environment due to the space-saving construction and the sophisticated design.

>> RO permeate output: 80 - 500 l/h
>> Pure water quality: 5 - 15 µS/cm
>> W/D/H: 1000/600/1900 mm


  1. Ready-to-connect deionized water treatment system – Plug&Treat
  2. Fully equipped closet system
  3. compact, high-quality deionized water treatment as RO system
  4. Completely closed, low-noise cabinet system
  5. Visualization, operation and parameterization via controller
  6. all necessary pre-treatment and post-treatment components integrated
  7. Connection to control technology possible
  8. Electrical and control expansion up to complete building supply with deionized water
  9. Measured value processing completely via PLC
  10. all measuring points can be calibrated and adjusted
  11. all conductivity displays are temperature-compensated
  12. horizontal non-wearing stainless steel inline high-pressure pump
  13. high-quality electrical and hydraulic brand components
  1. Quantity-controlled individual softening with chlorine electrolysis
  2. soft water monitoring
  3. Salt deficiency alarm
  4. Storage tank with sloping bottom
  5. electrodeionization
  1. Permeate Output: 80 - 500 l/h
  2. Pure Water Quality: 5 - 15 μS/cm (depending on the Raw Water Quality)
  3. Single Softener: 60 m³/°dH or 100 m³/°dH
  4. Brine Tank: 50 l
  5. Tank Volume (intern): 80 l
  6. RO-Fine Filter: 10“ or 20“
  7. Water inlet: DN 25
  8. Water outlet DN 25
  9. Drain: DN 50
  10. Connection Voltage: 230 V/50 Hz
  11. Raw Water Pressure: 3,0 - 6,0 bar
  12. Raw Water Temperature: 5°C - 25°C
  13. Ambient Temperature: 5°C - 30°C
  14. Dimensions in mm (W x D x H): 1000 x 600 x 1900
  15. Weight: approx. 450 kg, depending on the Version

Request more information about our HA-RO constrict

All fields marked with * are required for further processing of your request and must be filled out.

I agree that my details from the contact form will be collected and processed to answer my request. The data will be deleted after your request has been processed. You can revoke your consent at any time for the future by sending an email to info@hartmann-gmbh.eu. Detailed information on handling user data can be found in our Privacy Policy.